The Module determines CO2 concentration in the breathing gases by measuring the amount of light absorbed by these gases. The Module contains a reference channel, used to compensate for optical changes, so that the system is in a state of calibration. EtCO2 is displayed as a numerical value in millimeters of mercury (mmHg), percent (%), or kilopascals (kPa). In addition, a CO2 waveform (capnograph) may be displayed which is a valuable clinical tool that can be used to assess patient airway integrity and proper endotracheal tube (ETT) placement.
Respiration rate is calculated by measuring the time interval between detected breaths.
- Portable and Modular design : Elegant in appearance, clear in labeling and easy to operate. Modular design, small and lightweight. Work with monitor to continuously measure CO2 concentration in gas and calculate End-tidal CO2 (ETCO2), Inspired Minimum CO2 (InsCO2) and Airway Respiration Rate (AwRR)
- Optional measurement units: Multiple measurement units to choose: mmHg, kPa and %.
-
High spec measurement rangeCO2 measurement range: 0 to 150 mmHg
- CO2 measurement accuracy:
- 0~40 mmHg: ±2mmHg
- 41~70 mmHg: ±5% of reading
- 71~100 mmHg: ±8% of reading
- 101~150 mmHg: ±10% of reading
- CO2 measurement accuracy:
-
- Respiratory monitoring range:
- CO2-M02:2rpm~150rpm
- accuracy: ±1rpm
- Respiratory monitoring range: